Certificate No.:
IECEx SIR 08.0028X issue No.:3
 
Status:
Current
 
Date of Issue:2012-01-30 Page 1 of 5


    
Applicant:
Hydramotion Limited
New York House
1 York Road Industrial Estate
Malton
North Yorkshire YO17 6YA
United Kingdom


 
Electrical Apparatus:XL7 (aka ‘Viscojet’), ViscoMelt, Indie (aka ‘Reactavisc’, ‘Viscolite’ or ‘Go Visc’)
Optional accessory:


 
Type of Protection:Intrinsic Safety


 
Marking:

Ga Ex ia IIC T4 (-40°C to +80°C)
Ga Ex ia IIC T6 (-40°C to +55°C)

 
Approved for issue on behalf of the IECEx
Certification Body:
C Ellaby

 
Position:Deputy Certification Manager


 
Signature:
(for printed version)
 
Date:
 
 
1. This certificate and schedule may only be reproduced in full.
2. This certificate is not transferable and remains the property of the issuing body.
3. The Status and authenticity of this certificate may be verified by visiting the Official IECEx Website.
 

 Certificate issued by:
SIRA Certification Service
Rake Lane
Eccleston
Chester
CH4 9JN
United Kingdom


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx SIR 08.0028X
 
Date of Issue:2012-01-30Issue No.: 3
 
Page 2 of 5
 
 Manufacturer:
Hydramotion Limited
New York House
1 York Road Industrial Estate
Malton
North Yorkshire YO17 6YA
United Kingdom


 
 Additional Manufacturing location(s):
 

This certificate is issued as verification that a sample(s), representative of production, was assessed and tested and found to comply with the IEC Standard list below and that the manufacturer's quality system, relating to the Ex products covered by this certificate, was assessed and found to comply with the IECEx Quality system requirements. This certificate is granted subject to the conditions as set out in IECEx Scheme Rules, IECEx 02 and Operational Documents as amended.


STANDARDS:
The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents, was found to comply with the following standards:

 
IEC 60079-0 : 2004
Edition: 4.0
Electrical apparatus for explosive gas atmospheres - Part 0: General requirements
IEC 60079-11 : 2006
Edition: 5
Explosive atmospheres - Part 11: Equipment protection by intrinsic safety "i"
IEC 60079-26 : 2006
Edition: 2
Explosive atmospheres - Part 26: Equipment with equipment protection level (EPL) Ga
 
This Certificate does not indicate compliance with electrical safety and performance requirements other than those expressly included in the Standards listed above.


 
TEST & ASSESSMENT REPORTS:A sample(s) of the equipment listed has successfully met the examination and test requirements as recorded in

 
Test Report:
GB/SIR/ExTR08.0058/00GB/SIR/ExTR09.0038/00GB/SIR/ExTR09.0093/00GB/SIR/ExTR12.0017/00


Quality Assessment Report:

GB/SIR/QAR08.0013/00GB/SIR/QAR08.0013/01GB/SIR/QAR08.0013/02GB/SIR/QAR08.0013/03GB/SIR/QAR08.0013/04


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx SIR 08.0028X
 
Date of Issue:2012-01-30Issue No.: 3
 
Page 3 of 5
 
 
Schedule
 
EQUIPMENT:
Equipment and systems covered by this certificate are as follows:
 

There three types of viscometer, the XL7 (aka ‘Viscojet’), the ViscoMelt and the Indie (aka ‘Reactavisc’, ‘Viscolite’ or ‘Go Visc’).  All the instruments measure fluid viscosity by electrically sensing the damping, due to the fluid viscosity, experienced by a mechanical, torsionally resonating shaft system, part of which is immersed in the process fluid.  Two electromagnets connected to an amplifier module maintain the vibrational resonance of the shaft system.

 
CONDITIONS OF CERTIFICATION: YES as shown below:
 

1.

Some versions of this equipment use an aluminium alloy enclosure and ignition sources due to impact and friction sparks could occur.  This shall be considered when an aluminium alloy enclosure is installed.
2.The Indie, Reactavisc, Viscolite and Go Visc versions of this equipment may use a plastic enclosure that could generate an ignition-capable level of electrostatic charge under certain extreme circumstances.  Therefore, the equipment shall not be installed in a location where the external conditions may lead to the build-up of electrostatic charge on such surfaces, particularly in zone 0 applications.  Additionally, the equipment shall only be cleaned with a damp cloth.
 


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx SIR 08.0028X
 
Date of Issue:2012-01-30Issue No.: 3
 
Page 4 of 5
 
 
EQUIPMENT(continued):
 

XL7 Viscometer (aka the ViscoJet) incorporates a microcontroller in the smart head module allowing a bi-directional serial data link to safe area equipment. Frequency measurement is performed within the module. The transducer calibration constants are stored within smart head module, which can calculate viscosity. The smart head module has a platinum resistance thermometer (PRT) input allowing transducers equipped with a PRT to measure temperature. Viscosity, temperature and other parameters can be telemetered to safe area equipment via the serial data link. XL7 transducers can be supplied in a variety of different sizes, with many different process fittings. The only user connection is a 3-wire supply connection, which is by means of a 3-way terminal block inside the enclosure, the cable exiting the enclosure via a gland.

ViscoMelt Viscometer, this model is designed for high temperature and pressure applications and can be retrofitted to polymer extruders via a standard 1/2"-20 UNF threaded port, commonly used by temperature and pressure sensors. The only user connection is a 3-wire supply connection, which is by means of a 3-way terminal block inside the enclosure, the cable exiting the enclosure via a gland.
Indie Viscometer (aka Reactavisc, Viscolite and Go Visc) is a compact viscometer that uses the smart head module and has a stainless steel, aluminium alloy or plastic body.  In these instruments, the dimensions of the mechanically resonating sensor element vary between models, but in all cases the welded sensor forms a metallic barrier between process fluid and the instrument electronics.  The process connection/seal is an elastomeric compression-type gland, fitted over the welded resonator system.  This fitting is deliberately separated from the environmental seal of the electronics enclosure.  These instruments are usually equipped with an integral temperature sensor.  Field connections enter the plastic enclosure via a plug and socket.  The only user connection is by means of a plug-and-socket connector mounted on the wall of the enclosure.
All models have the following safety description:
Ui = 28 V dc, Ii = 0.12 A, Pi = 0.84 W, Ci = 0, Li = 10 μH
The Manufacturer shall note the following conditions of manufacture:
1.For viscometers designed for use at more than 10% above normal atmospheric pressure, the maximum design pressure rating shall be indicated in the documentation supplied to the user and the manufacturer shall carry out a routine pressure test on the wetted components and welds that are exposed to process pressure at 1.5 times the design pressure
2.The manufacturer shall test 100% of all manufactured units in accordance with the dielectric strength test in IEC 60079-11:2006 clauses 6.3.12 and 10.3. 500 Vrms shall be applied between the circuit and the metallic enclosure for a minimum of 60 s; the maximum current flowing shall not exceed 5 mA and there shall be no flashover or breakdown.
 
 
 


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx SIR 08.0028X
 
Date of Issue:2012-01-30Issue No.: 3
 
Page 5 of 5
 
 
DETAILS OF CERTIFICATE CHANGES (for issues 1 and above):
 

Issue 1 – this Issue introduced the following changes:

1.The addition of drawings detailing the positioning and wiring connections of the integral PRT temperature sensor in the XL7 viscometer and the Viscomelt polymer melt viscometer.
2.The inclusion of an additional Condition of Manufacture.
Issue 2 – this Issue introduced the following change:
1.A more generic definition of the coil cable was recognised, this is defined on the assembly drawing H21-04-062 Revision 6.
Issue 3 – this Issue introduced the following changes:
1Mechanical changes to the ‘Indie’ version to allow a higher process pressure were endorsed, the description was modified to reflect this change.
2A correction to the certificate holders’ postcode from YO17 0NW to YO16 6YA was approved.
 
 








Copyright © IEC-IECEx 2013 , Geneva Switzerland. All rights reserved.