IECEx CoC Home > Certified Equipment Scheme > Certificate of Conformity
important information !
an IECEx Guest User | ExCB Login

IECEx Certificate
of Conformity

 
INTERNATIONAL ELECTROTECHNICAL COMMISSION
IEC Certification Scheme for Explosive Atmospheres
for rules and details of the IECEx Scheme visit www.iecex.com

 
Certificate No.:
IECEx BAS 11.0140X issue No.:0

Certificate history:

 
Status:
Current
 
Date of Issue:2012-04-11 Page 1 of 3


    
Applicant:
GE Infrastructure Sensing, Inc.
1100 Technology Park Drive
Billerica
Massachusetts 01821
United States of America


 
Electrical Apparatus:Delta F Oxygen Cell
Optional accessory:


 
Type of Protection:Intrinsic Safety


 
Marking: Ex ia IIC T5 (-20°C ≤ Ta ≤ +50°C
 
Approved for issue on behalf of the IECEx
Certification Body:
R S Sinclair

 
Position:General Manager


 
Signature:
(for printed version)
 
Date:
 
 
1. This certificate and schedule may only be reproduced in full.
2. This certificate is not transferable and remains the property of the issuing body.
3. The Status and authenticity of this certificate may be verified by visiting the Official IECEx Website.
 

 Certificate issued by:
Baseefa
Rockhead Business Park
Staden Lane
Buxton
Derbyshire
SK17 9RZ
United Kingdom


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx BAS 11.0140X
 
Date of Issue:2012-04-11Issue No.: 0
 
Page 2 of 3
 
 Manufacturer:
GE Infrastructure Sensing, Inc.
1100 Technology Park Drive
Billerica
Massachusetts 01821
United States of America


 
 Additional Manufacturing location(s):
GE Sensing EMEA
Unit A/B
Shannon Free Zone East
Shannon
County Clare
Ireland
 

This certificate is issued as verification that a sample(s), representative of production, was assessed and tested and found to comply with the IEC Standard list below and that the manufacturer's quality system, relating to the Ex products covered by this certificate, was assessed and found to comply with the IECEx Quality system requirements. This certificate is granted subject to the conditions as set out in IECEx Scheme Rules, IECEx 02 and Operational Documents as amended.


STANDARDS:
The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents, was found to comply with the following standards:

 
IEC 60079-0 : 2011
Edition: 6.0
Explosive atmospheres - Part 0: General requirements
IEC 60079-11 : 2011
Edition: 6.0
Explosive atmospheres - Part 11: Equipment protection by intrinsic safety "i"
 
This Certificate does not indicate compliance with electrical safety and performance requirements other than those expressly included in the Standards listed above.


 
TEST & ASSESSMENT REPORTS:A sample(s) of the equipment listed has successfully met the examination and test requirements as recorded in

 
Test Report:
GB/BAS/ExTR11.0283/00


Quality Assessment Report:

GB/BAS/QAR06.0025/03GB/BAS/QAR10.0026/01


- PAGE BREAK -

IECEx Certificate
of Conformity

 
Certificate No.:IECEx BAS 11.0140X
 
Date of Issue:2012-04-11Issue No.: 0
 
Page 3 of 3
 
 
Schedule
 
EQUIPMENT:
Equipment and systems covered by this certificate are as follows:
 

The Delta F Oxygen Cell is designed to provide an output voltage proportional to the oxygen level present in local atmosphere. An optional temperature sensor provides an output current proportional to the absolute temperature.

The oxygen sensor comprises two "stab-el" electrodes and two sensor electrodes all housed in a plastic and metallic enclosure. External connections are made via terminals fitted in the wall of the sensor. The optional temperature sensor is mounted in a recess in the enclosure. External connections are made via a short fixed cable and a flying connector.

Oxygen Channel Terminal Parameters
Ui = 23V
Ii  = 75mA
Pi = 450mW
Ci = 0
Li = 0

Temperature Channel Terminal Parameters
Ui = 23V
Ii  = 15mA
Pi = 90mW
Ci = 0
Li
= 0

 
CONDITIONS OF CERTIFICATION: YES as shown below:
 

1. The apparatus is to be installed such that the apparatus terminals are protected to at least the requirements of IP20.
2. The sensor must not be used to measure oxygen concentrations greater than 21%
3. To minimise the risk from electrostatic discharge the plastic body should only be cleaned with a damp cloth.
4. The gas connectors are electrically isolated and when considered in parallel they have a capacitance of 10pF. This must be taken into account when mounting the equipment.

 